Top class impedance measurement and signal generation - All in one device. SIGGI combines all functions required for
objectively verifying any device for acquisition, amplification and processing
of electrophysiological signals (EEG, ECG, EMG, etc.) in SIGGI reliably and quickly ensures the validity SIGGI is a must for any electrophysiological laboratory or medical technology department. |
|
|
OPERATION |
|
All functions of SIGGI advanced
and SIGGI basic
can be reached intuitively via the functions menu and the four navigation
keys.
|
![]() |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
CONNECTORS |
SignalsSignal generator ground is connected to output A. Connect this output to the reference input of your device, e.g. of your EEG amplifier, and shortcut it with Gnd. The generated test signal is output to connector C. Connect this lead to any desired channel of your equipment. Connector B delivers the generated signal at half the original signal amplitude found at connector C.
|
![]() |
|
TriggerAt the start of every signal period, a TTL trigger signal of half period
duration is output to the trigger connector. Connect the trigger output
with the corresponding input of your device (e.g. pins 1 and 14 on the
PC-card trigger input of the BrainAmp®, pins 8 and 25 on the
trigger input connector of the SynAmp®). You may have to change
trigger polarity in the software setup. |
![]() |
ImpedanceImpedance is measured between Impedance-In und Impedance-Out. Connect your electrode pair to these two connectors. Note: While it is possible to use rectangle signals for impedance
measurement with SIGGI, we do not recommend
this, as this measurement procedure is mathematically undefined. |
![]() |
|
|
SOFTWARE |
|
SIGGI advanced comes with the Windows program SIGGI.EXE, which can be used to display and change any menu parameter available in SIGGI advanced directly on your PC. Additionally, you can use SIGGI.EXE to transfer any test signal you desire
into the signal memory of SIGGI advanced,
which can then be used for testing your equipment. Of course, the use
of the built in sine, rectangle and DC signals for equipment tests is
not affected by a loaded user signal. - SIGGI.EXE can read the most common
data arrangements in ASCII format. - Examples can be found in the folder
\EXAMPLES. |
||
|
Installation: Deinstallation: |
||
|
|
POWER SUPPLY |
|
SIGGI is powered by rechargable batteries. Typical operating time is approximately 10 hours with a recharging time of about 14 hours. - After six minutes of operation without a key press SIGGI switches off automatically. |
![]() |
|
| Red diode indicates low or empty batteries. Green diode indicates battery charge operation. |
![]() |
|
|
|
Subject and patient safety |
|
In order to connect the battery charger or the serial cable used for communication with the PC, the mechanical slider has to be dislocated.This effectively blocks the signal and impedance outputs and thus blocks operation of the device. This reliably prevents a subject or patient from accidently being connected to the mains power and suffering injuries in case of an unexpected failure of the mains adapter or of the computer connected to SIGGI. |
|
|
DELIVERY SCOPE |
|
|||||||||||||
|
|
TECHNICAL SPECIFICATIONS |
Impedance measurement
Signal generatorSine and Square wave signal
DC-Signal Common mode rejection test User defined signal Trigger output Internal resistance Impedance Measurement Calibration
|